Accessory Pack for Spectroscopy
Orthogonal Signal Correction enables the creation of improved models when spectral data contains a great deal of interference or random variations unrelated to the calibration or prediction.

Calibration Transfer functions save time by allowing calibration of instruments using fewer calibration samples. This function also facilitates the development of more robust models reducing the need to recalibrate.

Deresolv can be used to change the apparent resolution of an instrument.
 
Orthogonal Signal Correction (OSC)

This function calculates the orthogonal signal correction parameters for preprocessing a spectral data prior to developing a PLS model. In situations where the PLS model captures a very large amount of predictor block (X) variance in the first factor but gets very little of the predicted variable (y or Y) it can be very useful. In these cases it is useful to remove extraneous variance from X that is unrelated to y. OSC does this by finding directions in X that describe large amounts of variance while being orthogonal to y. Models are built on OSC corrected data, then the OSC correction is applied before application of the PLS models to new data.


NIR spectra of different alcohols

Deresolv (high/low resolution instrument)

On occasion, one may wish to standardize a lower resolution instrument to a higher resolution instrument. In such an instance, it may be more effective to mathematically lower the resolution of the higher resolution instrument prior to forming the transfer model. The Deresolv function can be used to change the apparent resolution of an instrument and may also generate some beneficial noise reduction. Deresolv works by FFTing the spectra, convolving it with an impulse of specified width, then back FFTing the result.

Calibration Transfer Functions, (sample selection to do standardization)

The relationship between the responses of the two instruments is obtained using spectra from a set of standardization samples or using a subset of samples from the Instrument 1 calibration set. Selection of this subset can be performed with a function. The function selects samples based on their leverage, or uniqueness, within the group of spectra. Generally, 3 spectra are sufficient for building a transfer function between instruments.

The Unscrambler® Accessory Pack for Spectroscopy is developed in collaboration with

CAMO continues its commitment to bring you today's most powerful modeling advancements and innovations. Our first The Unscrambler® Accessory Pack not only shows the flexibility and strength of the new features adopted into The Unscrambler® but also demonstrates CAMO's ability to understand, respond to, and satisfy customer needs more rapidly.